Frequently Asked Questions (FAQs)
1. What is the OSHA standard for fume hoods?
According to OSHA's Laboratory standard (29 CFR 1910.1450), fume hoods in laboratories must be properly maintained and functioning. Users should receive training on how to use the fume hood correctly to ensure safety.
The EPA Performance Requirements for Laboratory Fume Hoods specify criteria for chemical fume hoods. These requirements include average face velocities of 100 fpm for typical chemical fume hoods and 60 fpm for low-velocity chemical fume hoods.
3. What are the standards for laboratory hoods?
Standards for laboratory hoods include ensuring general airflow is uniform with no turbulent or static areas, and face velocity should be adequate (typically 60-100 lfm). The airflow into and within the hood should not be excessively turbulent.
4. What are the rules for fume hoods in a laboratory?
When working with fume hoods in a laboratory, it is important to wear safety glasses (goggles), lab coats, and appropriate chemical gloves. Additionally, do not put your head beyond the sash opening of the hood, avoid using the hood for storage, and follow proper safety guidelines.
5. What is required by the OSHA lab standard?
The OSHA Laboratory standard includes five major elements: hazard identification, chemical hygiene plan, information and training, exposure monitoring, and medical consultation and examinations.
6. How do I choose a fume hood?
When choosing a fume hood, consider factors such as the size of the room, the number of room air changes, lab heat load, types of materials used, and the required linear feet of hood based on your needs.
7. How many CFM do I need for a chemical fume hood?
The required air volume passing through a fume hood depends on the area of the sash opening and the desired average velocity. For example, if a velocity of 100 fpm is needed and the hood has a sash opening of 7.5 square feet, the air volume required would be 750 CFM.
8. What are the safety guidelines for fume hoods?
To ensure safety when using fume hoods, it is important to conduct work at least 6 inches back from the face of the hood, keep the sash closed as much as possible, ensure hood slots and baffles are not obstructed, and avoid storing apparatus or chemicals permanently in the hood.
9. What are the air flow requirements for a lab fume hood?
Laboratory fume hoods should provide a minimum average effective face velocity of 100 fpm, with a minimum of 70 fpm at any point.
10. What are four guidelines to remember for using fume hoods properly?
To use fume hoods properly, remember to conduct work at least 6 inches back from the hood's face, keep the sash closed as much as possible, ensure hood slots and baffles are not obstructed, and avoid storing apparatus or chemicals permanently in the hood.